In vivo heating of pacemaker leads during magnetic resonance imaging

期刊

EUROPEAN HEART JOURNAL
卷 26, 期 4, 页码 376-383

出版社

OXFORD UNIV PRESS
DOI: 10.1093/eurheartj/ehi009

关键词

magnetic resonance imaging; pacemakers; heating; in vivo; safety; electromagnetic interference

向作者/读者索取更多资源

Aims Magnetic resonance imaging (MRI) is well established as an important diagnostic tool. in medicine. However, the presence of a cardiac pacemaker is usually regarded as a contraindication for MRI due to safety reasons. In this study, heating effects at the myocardium-pacemaker lead tip interface have been investigated in a chronic animal model during MRI at 1.5 Tesla. Methods and results Pacemaker leads with additional thermocouple wires as temperature sensors were implanted in nine animals. Temperature increases of up to 20 degrees C were measured during MRI of the heart. Significant impedance and minor stimulation threshold changes could be seen. However, pathology and histology could not clearly demonstrate heat-induced damage. Conclusions MRI may produce considerable heating at the lead tip. Changes of pacing parameters due to MRI could be seen in chronic experiments. Potential risk of tissue damage cannot be excluded even though no reproducible alterations at the histological level could be found.