期刊
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 52, 期 2, 页码 153-159出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2004.840188
关键词
fetal surveillance; Monte Carlo simulation; reflectance pulse oximetry; tissue optics
A multilayer tissue description was employed in Monte Carlo simulations of reflectance pulse oximetry to study the impact of assumptions made in previous studies employing homogeneous tissue models. Simulation results with a discrete layer of arterial pulsatility were, similar to previous studies employing homogenous tissue models. However, the relationship of normalized pulse amplitude to emitter-detector spacing reiterates that spacing has a significant impact on pulse oximetry function. The effect of melanin, content as a thin, superficial absorber was also simulated, with results supporting the general clinical observation that skin shade need not substantially compromise pulse oximeter accuracy.