4.7 Article

Microfabrication and test of a three-dimensional polymer hydro-focusing unit for flow cytometry applications

期刊

SENSORS AND ACTUATORS A-PHYSICAL
卷 118, 期 2, 页码 259-267

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.sna.2004.09.001

关键词

SU-8; three-dimensional hydro-focusing; microfluidic; microchannel; cytometer

向作者/读者索取更多资源

This paper details a novel true three-dimensional (3D) hydro-focusing microcell sorter for micro-flow cytometry applications. The unit was microfabricated by means of SU-83D lithography. The 3D microstructure for coaxial sheathing was designed, microfabricated, and tested. Three-dimensional hydro-focusing capability was demonstrated with an experiment to sort labeled tanned sheep erythrocytes (red blood cells). This polymer hydro-focusing microstructure is easily microfabraicted and integrated with other polymer microfluidic structures. (C) 2004 Elsevier B.V. All rights reserved.