4.8 Article

Instrumentation and method for ultrahigh resolution field desorption ionization Fourier transform ion cyclotron resonance mass spectrometry of nonpolar species

期刊

ANALYTICAL CHEMISTRY
卷 77, 期 5, 页码 1317-1324

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac048766v

关键词

-

向作者/读者索取更多资源

We describe the construction and application of a 9.4-T Fr-ICR mass spectrometer interfaced to a commercial field desorption ion source for high-resolution, high-mass accuracy measurements of nonpolar species. The Fr-ICR MS instrument includes a liquid injection field desorption ionization source, octopole ion guides, external octopole ion trap capable of an axial potential gradient for ion ejection, capacitively coupled open cylindrical ion trap, and pulsed gas valve for ion cooling. Model compound responses with regard to various source and instrument conditions provide a basis for interpretation of broadband mass spectra of complex mixtures. As an example, we demonstrate broadband speciation of a Gulf Coast crude oil, with respect to numerous heteroatomic classes, compound types (rings plus double bonds), and carbon number distributions.