4.7 Article

LC/MS determination of bisphenol A in river water using a surface-modified molecularly-imprinted polymer as an on-line pretreatment device

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 381, 期 6, 页码 1193-1198

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-004-3031-1

关键词

water analysis; molecular imprinting; column switching; endocrine disruptor; bisphenol A; pretreatment; LC/MS

向作者/读者索取更多资源

A new analytical method for the determination of bisphenol A (BPA) by LC/MS was developed and applied to environmental water samples. Quantitative MS detection of BPA was carried out in the negative mode. In order to preconcentrate the target compound yet prevent serious contamination of the water samples from the experimental environment, we employed column switching HPLC coupled with a pretreatment column of surface-modified molecularly-imprinted polymers. The recovery of BPA from a spiked environmental water sample was 102% and the repeatability of actual determinations of water samples containing 20 ng/ L of BPA was 5.4% RSD. By modifying the surfaces of the molecularly-imprinted polymer particles packed in the pretreatment column, interference from the water samples was effectively removed, resulting in a significant increase in sensitivity and more reliable results. This method was successfully applied to the trace determination of BPA in environmental water samples using LC/ MS.