4.4 Article Proceedings Paper

Electrochemical behavior of diamond-like carbon films for biomedical applications

期刊

THIN SOLID FILMS
卷 475, 期 1-2, 页码 291-297

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.tsf.2004.07.052

关键词

DLC; porosity; protective efficiency; SEM; EIS

向作者/读者索取更多资源

Diamond-like carbon (DLC) films have several advantages in biomedical applications such as high hardness, chemical inertness, low fiction and electrical insulation. Furthermore, DLC-coated STS 316L films have been reported to have a good biocompatibility. In addition, corrosion resistance is the first consideration for the biomaterials to be used in the body. DLC films have been deposited onto substrates of STS 316L using rf plasma-assisted chemical vapor deposition (PACVD) with benzene (C6H6) or a mixture Of C6H6 and silane (SiH4) as process gas. Three kinds of DLC films were prepared as a function of bias voltage and Si incorporation. Corrosion behavior of DLC films was investigated by electrochemical techniques (potentiodynamic polarization test and electrochemical impedance spectroscopy (EIS)) and surface analytical techniques (atomic force microscopy (AFM) and scanning electron microscopy (SEM)). The electrolyte used in this test was a 0.89% NaCl of pH 7.4 at temperature of 37degreesC. Electrochemical measurements showed that DLC films with higher bias voltage and Si incorporation could improve corrosion resistance in the simulated body fluid environment. (C) 2004 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据