期刊
ELECTROPHORESIS
卷 26, 期 7-8, 页码 1576-1583出版社
WILEY
DOI: 10.1002/elps.200410058
关键词
capillary zone electrophoresis; electrospray ionization-mass spectrometry; melamine resins; time of flight-mass spectrometry
A method for the determination of the major components of (methoxymethyl)melamine resins, with quantitative analysis of unreacted melamine by capillary zone electrophoresis (CZE) using electrospray ionization-mass spectrometry (ESI-MS) is presented. Using a low background electrolyte (BGE) pH, components are separated according to their charge/ionic radius ratio with a distinctly different separation selectivity compared to the HPLC methods commonly employed in melamine-resin analysis. The use of a time-of-flight mass spectrometer (TOF-MS) was concluded to be necessary, as the complex samples studied required maximum sensitivity and resolution, which is clearly superior for TOF-MS detectors over their quadrupole counterparts. A standard curve of free melamine was determined with an R-2 = 0.999 over a concentration range of an order of magnitude. This method offers the unique separation selectivity of CZE as well as a quicker analysis time, especially for dimers compared to the HPLC methods used to date.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐