期刊
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 28, 期 4, 页码 270-273出版社
WILEY
DOI: 10.1111/j.1540-8159.2005.09520.x
关键词
magnetic resonance imaging; ICD; electromagnetic interference; safety
Objective: To determine if simple strategies used to safely scan pacemaker patients could be applied to implantable cardioverter defibrillator (ICD), patients undergoing MRI allowing ICD patients to undergo MRI as well. Interventions: Screening, reprogramming, and monitoring strategies were used to facilitate MRI. Results: Seven patients underwent eight MRI scans at 1.5 T Post-MRI, all devices demonstrated no change in pacing, sensing, impedances, charge times, or battery status. The patient undergoing a lumbar spine scan experienced a power-on-reset of his ICD without permanent impairment of his device. Conclusion: Scanning of ICD patients might be performed if appropriate reprogramming and monitoring is implemented.
作者