4.6 Article

Robot-assisted biopsy using computed tomography-guidance -: Initial results from in vitro tests

期刊

INVESTIGATIVE RADIOLOGY
卷 40, 期 4, 页码 219-228

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/01.rli.0000155285.05672.cf

关键词

medical robotics; interventional radiology; surgical robotics; biopsies; computed tomography (CT)

向作者/读者索取更多资源

Purpose: We sought to develop a robotic system for computed tomography (CT)-guided biopsy to validate the feasibility, accuracy, and efficacy of the system using phantom tests. Materials and Methods: Ten peas (mean diameter 9.9 +/- 0.4 mm) embedded within a gel phantom were selected for biopsy. Once the best access was defined on CT images, the position of the phantom was recorded by an optical tracking system. Positional data about the phantom and the corresponding CT image was transferred to the robot planning system (Linux-based industrial PC equipped with video capture card). Once the appropriate position, angulation, and pitch were calculated, the robotic arm moved automatically with 7 degrees-of-freedom to the planned insertion path, aiming the needle-trajectory at the center of the target. Then, the biopsy was performed manually using a coaxial technique. The length of all harvested specimens was measured and short cut pieces of a guidewire were pushed into the target to evaluate the deviation of the actual needle track from the target. Results: In all targets, biopsy specimens (mean length 5.6 +/- 1.4 mm) were harvested with only 1 needle pass required. The mean deviation of the needle tip from the center of the target in the x and z axes was 1.2 +/- 0.9 turn and 0.6 +/- 0.4 mm, respectively. Conclusions: Robotic-assisted biopsies in vitro, using CT guidance, were feasible and provided high accuracy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据