期刊
ELECTROANALYSIS
卷 17, 期 8, 页码 648-654出版社
WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200403134
关键词
cadmium ion selective electrode; potentiometric sensor; PVC membrane; Schiff base
5-[((4-Methyl phenyl) azo)-N-(6-amino-2-pyridin) salicylaldimine] (S-1), and 5-[((4-methyl phenyl) azo)-N-(2-diamino-2-cyano-1-ethyl cyanide) salicylaldehyde] (S-2) with N and O donor atoms are effective ionophores to make Cd2+-selective membrane electrodes. The electrodes based on S-1 and S-2 exhibits a Nernstian or near-Nernstian response for cadmium ion over a wide concentration range 1.5 x 10(-1)-7.5 x 10(-7) with a slope of 28 and 2.0 x 10(-1)-4.0 x 10(-7) with a slope of 22, respectively. They have quick response and can be used for three or four months without any divergence in potential. The proposed sensors show fairly good selectivity over some alkali, alkaline earth, transition and heavy metal ions. The electrodes based on S-1 and S-2 can be used in the pH range 3.5-9. These electrodes were used as an indicator electrode in potentiometric titration of cadmium ion with EDTA and in the direct determination of cadmium ion in aqueous solutions.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据