4.6 Article

Imaging magnetically labeled cells with magnetomotive optical coherence tomography

期刊

OPTICS LETTERS
卷 30, 期 7, 页码 747-749

出版社

OPTICAL SOC AMER
DOI: 10.1364/OL.30.000747

关键词

-

类别

向作者/读者索取更多资源

We introduce a novel contrast mechanism for optical coherence tomography (OCT) whereby the optical scattering of magnetically labeled cells is modified by means of an externally applied magnetic field. This modification is made through the addition of a small electromagnet to the imaging arm of a conventional OCT interferometer. We measure the magnetomotive OCT signal by differencing pairs of axial scans (A-scans) acquired with the magnetic field on and off. Magnetomotive contrast is demonstrated in bulk three-dimensional cell scaffolds containing macrophages labeled with microparticles of iron oxide, demonstrating magnetic-specific contrast over a dynamic range of 30 dB. (c) 2005 Optical Society of America.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据