4.3 Article

A comparison of plates with and without locking screws in a calcaneal fracture model

期刊

FOOT & ANKLE INTERNATIONAL
卷 26, 期 4, 页码 309-319

出版社

AMER ORTHOPAEDIC FOOT & ANKLE SOC, INC
DOI: 10.1177/107110070502600407

关键词

calcaneal fracture; experimental testing; fracture model; locked screw; plate fixation

向作者/读者索取更多资源

Background: We compared different plates in an experimental calcaneal fracture model under biocompatible loading. Methods: Four plates were tested: a plate without locked screws (Synthes), and three different plates with locked screws (Newdeal, Darco, Synthes). Synthetic calcanei (Sawbone) were osteotomized to create a fracture model, and the plates were fixed onto them. Seven specimens for each plate model were subjected to cyclic loading (preload 20 N, 1,000 cycles with 800 N, 0.75 mm/s), and load to failure (0.75 mm/s). Motion, forces, plastic deformation of the plate, and consequent depression of the posterior joint facet were analyzed. Results: During cyclic loading, all plates with locked screws showed statistically significant lower displacement in the primary loading direction than the plates without locked screws. Mean values (mm) of maximal displacements for each plate during cyclic loading were as follows: Synthes, 3.5; Darco, 4.5; Newdeal, 5.0; Synthes without locked screws, 7.5; (p < 0.001). No statistically significant differences between the plates were found in relation to loads to failure and corresponding displacement. Conclusion: This is the first biomechanical study to assess the stability of different plates currently in use in our practice for the fixation of calcaneal fractures. Our results showed that plates with locked screws provided greater stability during cyclic loading than the plate without locked screws.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据