We combine fiber Bragg grating (FBG) technology with a wet chemical etch-erosion procedure and demonstrate two types of refractive index sensors using single-mode optical fibers. The first index sensor device is an etch-eroded single FBG with a radius of 3 mu m, which is used to measure the indices of four different liquids. The second index sensor device is an etch-eroded fiber Fabry-Perot interferometer (FFPI) with a radius of similar to 1.5 mu m and is used to measure the refractive indices of isopropyl alcohol solutions of different concentrations. Due to its narrower resonance spectral feature, the FFPI sensor has a higher sensitivity than the FBG sensor and can detect an index variation of 1.4 X 10(-5). Since we can measure the reflection signal, these two types of sensors can be fabricated at the end of a fiber and used as point sensors. (C) 2005 American Institute of Physics.