4.8 Article

High-speed, high-resolution monolithic capillary LC-MALDI MS using an off-line continuous deposition interface for proteomic analysis

期刊

ANALYTICAL CHEMISTRY
卷 77, 期 8, 页码 2323-2331

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac048322z

关键词

-

资金

  1. NIGMS NIH HHS [GM15847] Funding Source: Medline

向作者/读者索取更多资源

High-speed, high-resolution LC separations, using a poly(styrene-divinylbenzene) monolithic column, have been coupled to MALDI MS and MS/MS through an off-line continuous deposition interface. The LC eluent was mixed with alpha-cyano-4-hydroxycinnamic acid matrix solution and deposited on a MALDI plate that had been precoated with nitrocellulose. Deposition at subatmospheric pressure (80 Torr) formed a 250-mu m-wide serpentine trace with uniform width and microcrystalline morphology. The deposited trace was then analyzed in the MS mode using a MALDI-TOF/TOF MS instrument. Continuous deposition allowed interrogation of the separation with a high data sampling rate in the chromatographic dimensions, thus preserving the high resolution of narrow peaks (3-5-s peak width at half-height) of the fast monolithic LC. No extracolumn band broadening due to the deposition process was observed. Over 2000 components were resolved in a 10-min linear gradient separation of the model sample, and 386 unique peptides were identified in the subsequent MS/MS analysis. The continuous deposition interface allows the coupling of high-resolution separations to MALDI MS without degradation in separation efficiency, thus enabling high-throughput proteome analysis.