4.5 Article

Effect of vessel curvature on Doppler derived velocity profiles and fluid flow

期刊

ULTRASOUND IN MEDICINE AND BIOLOGY
卷 31, 期 5, 页码 663-671

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2005.01.011

关键词

Doppler; velocity profiles; CFD; secondary velocity

向作者/读者索取更多资源

Side-branches and curvatures in the arterial tree yield deviations from the axial oriented velocity. Velocity or volume flow estimates based on the assumption that flow is axially oriented are of limited value at these sites. This article evaluates information obtainable by using a multigate Doppler ultrasound (US) instrument used with curved phantoms, which resemble the human coronary arteries. The comparison of experimental velocity data with data provided by an accurate computational fluid dynamics (CFD) method shows differences in the range of 4 to 11% for four curvatures with different radii. Multigate data are also used to estimate the volume flow in the curved segments at different experimental conditions. An error lower than 15% is obtained, to be compared with a 24% error obtained by assuming a parabolic velocity profile. In particular, it is shown that the residual error is not related to the small deviation of the velocity vectors from the axial direction due to the presence of secondary velocity components, which are found to be of magnitude less than 10% with respect to the axial velocity component.© 2005 World Federation for Ultrasound in Medicine & Biology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据