期刊
INTERNATIONAL JOURNAL OF AUDIOLOGY
卷 44, 期 6, 页码 317-320出版社
TAYLOR & FRANCIS LTD
DOI: 10.1080/14992020500057822
关键词
otoacoustic emission; middle ear; pressure
Middle ear pathology has a negative effect on the detectability of otoacoustic emissions. In this study, we investigated the effect of compensating a deviant static middle ear pressure while measuring transient evoked otoacoustic emissions (TEOAEs). In 59 children (mean age 4 years, 5 months) TEOAEs were measured twice in one session: first at ambient pressure and than at compensated middle ear pressure. On average, TEOAE amplitudes increased by 1.9 dB as a result of middle ear pressure compensation. The amplitude increase was largest in frequency bands centred at I and 2 kHz and a statistically significant correlation was found between the amount of compensated pressure and the TEOAE amplitude increase. In the higher frequency bands centred at 3 and 4 kHz, TEOAE amplitudes were almost insensitive to pressure compensation. These results show that measuring OAEs at compensated middle ear pressure enhances the amplitude of TEOAEs, and thus improves the detectability.