4.3 Article

Efficient compensation of low-frequency magnetic field disturbances in NMR with fluxgate sensors

期刊

JOURNAL OF MAGNETIC RESONANCE
卷 174, 期 2, 页码 287-291

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jmr.2005.02.018

关键词

compensation of magnetic field disturbances; fluxgate sensors; field-frequency lock; correlation spectroscopy; H-2 NMR

向作者/读者索取更多资源

A simple stabilization scheme of B-0 magnetic field fluctuations is described. The method is based on external measurements of time dependent magnetic field fluctuations by fluxgate sensors and generation of a compensating correction current in a coil mounted directly on an NMR magnet. It is shown that such an approach efficiently eliminates relatively slow magnetic field variations with frequency up to approximately 100 Hz. In combination with a standard 2 H field-frequency lock system, the method enables acquisition of reproducible lineshapes and dramatically improves overall performance of a high resolution NMR spectrometer. The presented solution might substitute for the internal lock system in these case where deuterium lock is not available. (c) 2005 Elsevier Inc. All rights reserved.