4.4 Article Proceedings Paper

High-Tc superconducting receiving coils for nuclear magnetic resonance imaging

期刊

IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY
卷 15, 期 2, 页码 1326-1329

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TASC.2005.849582

关键词

HTS RF surface coil; magnetic resonance imaging; signal-to-noise ratio

向作者/读者索取更多资源

Nuclear magnetic resonance (NMR) microscopy poses high demands on the sensitivity of the receiver coils. We have developed high-T-c superconducting (HTS) tape receiving coils for nuclear magnetic resonance imaging. The surface receiver coil is constructed from high-T-c Bi2Sr2Ca2Cu3Oy tape coil and cooled in liquid nitrogen temperature. The desired receiver surface coil is numerically simulated and optimized to have high value of the unloaded quality factor. With this HTS receiver coil we have obtained significant improvement in the unloaded Q-value, loaded Q-value and substantial gain in signal-to-noise ratio (SNR). The SNR improvement of 2.4 was achieved in the kiwi imaging and improvement of 2.7 in the braining imaging of rat at 300 K. The MRI microscopy is tested and compared the results with copper receiver.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据