4.7 Review

Spiral imaging: A critical appraisal

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 21, 期 6, 页码 657-668

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/jmri.20320

关键词

magnetic resonsance imaging; echo-planar imaging; single-shot imaging; spiral imaging; regridding

向作者/读者索取更多资源

In view of recent application in cardiovascular and functional brain imaging, this work revisits the basic performance characteristics of spiral imaging in direct comparison to echo-planar imaging (EPI) and to conventional rapid gradient-echo imaging. Using both computer simulation and experiments on phantoms and human subjects at 2.9 T, the study emphasizes single-shot applications and addresses the design of a suitable trajectory, the choice of a gridding algorithm, and the sensitivity to experimental inadequacies. As a general results, the combination of a spiral trajectory with regridding of the k-space data poses no principle obstacle for high-quality imaging. On the other hand, experimental difficulties such as gradient deviations, resonance offset contributions, and concomitant field effects cause more pronounced and even less acceptable image artifacts than usually obtained for EPI. Moreover, when ignoring parallel imaging strategies that are also applicable to EPI, improvements of image quality via reduced acquisition periods are only achievable by interleaved multishot spirals because partial Fourier sampling and rectangular fields of view (FOVs) cannot be exploited for non-Cartesian trajectories. Taken together, while spiral imaging may find its niche applications, most high-speed imaging needs are more easily served by EPI.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据