4.4 Article

Design and features of the acorn CorCap™ stop cardiac support device:: The concept of passive mechanical diastolic support

期刊

HEART FAILURE REVIEWS
卷 10, 期 2, 页码 101-107

出版社

SPRINGER
DOI: 10.1007/s10741-005-4637-x

关键词

CorCap (TM) supercript stop cardiac support device; heart failure; passive constraint; wall stress; compliance

向作者/读者索取更多资源

The Laplace equation points to the importance of ventricular wall stress as a factor in heart failure development and progression. Based on animal and clinical experience with active and passive cardiomyoplasty, a synthetic passive constraint was proposed as a means of reducing excessive wall stress, and thus assuaging disease progression. The Acorn CorCap (TM) stop CSD Cardiac Support Device was designed to provide passive diastolic support and serve as a constraint against chronic cardiac dilation in heart failure. Basic and uncomplicated in appearance, the device nevertheless incorporates numerous sophisticated structural and functional features intended maximize therapeutic value. The device is fabricated from medical grade multifilament polyester yarn, in a specific knit construction chosen to provide structural integrity, low physical profile, pliability for intimate contact with the epicardial surface, ease of manipulation during implantation, and anisotropic compliance characteristics to encourage beneficial reverse remodeling.