4.5 Article

Single particle laser ablation time-of-flight mass spectrometer: An introduction to SPLAT

期刊

AEROSOL SCIENCE AND TECHNOLOGY
卷 39, 期 6, 页码 554-568

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/027868291009242

关键词

-

向作者/读者索取更多资源

We present our single particle mass spectrometer we call SPLAT. SPLAT was designed to make possible the detection and characterization of particles down to 100 nm, generate reproducible single particle mass spectra, and collect data at a high rate. The final instrument presented here is capable of characterizing individual particles down to 50 nm at the rate of 20 particles per second. In SPLAT mass spectra are generated by a two-step process, that uses a pulsed infrared laser to heat the particle and a time delayed pulsed UV laser to create ions. We describe a mode of operation that makes it possible to take advantage of the gas phase ionization of the semivolatile components of the particle, while also generating mass spectral signatures of the nonvolatile fraction, thereby providing complete particle mass spectra. We present some sample results from two field deployments of SPLAT.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据