4.5 Article

Proficiency testing (external quality assessment) of drug detection in oral fluid

期刊

FORENSIC SCIENCE INTERNATIONAL
卷 150, 期 2-3, 页码 161-164

出版社

ELSEVIER IRELAND LTD
DOI: 10.1016/j.forsciint.2004.11.025

关键词

proficiency; oral fluid; amfetamines; cannabinoids; cocaine; opiates; human gamma-globulin

向作者/读者索取更多资源

Eighteen external quality assessment (proficiency testing) samples were prepared from client specimens collected with the Intercept (R) oral fluid collection device and by spiking drug-free oral fluid. Samples were circulated in pairs at quarterly intervals to 13 UK and USA based laboratories for analysis by a panel of OraSure micro-plate Intercept (R) enzyme immunoassay kits and hyphenated mass spectrophotometric techniques. During the survey, there was a single case of non-specificity in a false report for methadone. The major errors were of lack of sensitivity relative to the concentration thresholds specified for the immunoassays. The sensitivity for overall 'present'/'not found' reports calculated as true positives/(true positives + false negatives) were for the amfetamine specific assay 50%, methyl-amfetamines 93%, barbiturates 64%, cannabinoids 73%, cocaine and metabolites 100%, benzodiazepines 69%, methadone 95%, opiates 79% (opiates excluding oxycodone 93%), phencyclidine 93% and human gamma-globulin 97%. A small number of the sensitivity errors were attributable to errors in chromatographic confirmation techniques. (c) 2005 Elsevier Ireland Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据