4.7 Article

Estimation of the uncertainty associated with the standard addition methodology when a matrix effect is detected

期刊

TALANTA
卷 66, 期 5, 页码 1168-1174

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2005.01.021

关键词

standard addition methodology; uncertainty; metal chelates; liquid chromatography

向作者/读者索取更多资源

If a matrix effect is detected during the validation stage of an analytical method, the standard addition methodology must be applied. To reach a good estimation of the uncertainty associated with the determination, an adequate identification and evaluation of each uncertainty source should be done. As an example to illustrate how to calculate the uncertainty in this case, the simultaneous determination of V(V) and Mo(VI) at trace levels after precolumn chelation and extraction with N-benzoyl-N-phenylhydroxylamine, has been selected. The final results show that the main contributions to the relative overall uncertainty are those closely related with the chemical aspects of the method, i.e. liquid-liquid extraction and standard addition calibration. (c) 2005 Elsevier B.V. All rights reserved.