4.7 Article

Photometric detection in flow analysis systems using integrated PEDDs

期刊

TALANTA
卷 66, 期 5, 页码 1340-1344

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2005.01.054

关键词

LEDs; optical sensing; pH sensing; transmission measurements; colorimetric analysis

向作者/读者索取更多资源

A novel inexpensive optical-sensing technique has been developed for colorimetric flow analysis. This sensing system employs two LEDs whereby one is used as the light source and the other as a light detector. The LED used as light detector is reverse biased with a 5-V supply so that the photocurrent generated by the incident light discharges the capacitance. Direct digital output is provided by a simple timer circuit that measures the time taken for this discharge process from 5 V (logic 1) to 1.7 V (logic 0). This sensing concept has been applied in flow analysis by constructing an optical flow cell with a pair of LEDs. Calibration of the integrated optical flow cell using a dye resulted in a linear response that obeys the Beer-Lambert law. The flow rate, dynamic range, sensitivity and limits of detection were investigated. The system was also used for pH determination in the range of pH 2.5-6.8 using bromocresol green (BCG). The pK(a) of BCG was successfully determined by this technique. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据