4.7 Article

Influence of intracoronary attenuation on coronary plaque measurements using multislice computed tomography: observations in an ex vivo model of coronary computed tomography angiography

期刊

EUROPEAN RADIOLOGY
卷 15, 期 7, 页码 1426-1431

出版社

SPRINGER
DOI: 10.1007/s00330-005-2697-x

关键词

multislice computed tomography; coronary angiography; ex vivo experiment; vascular attenuation; coronary plaque attenuation

向作者/读者索取更多资源

Assessment of attenuation (measured in Hounsfield units, HU) of human coronary plaques was performed using multislice computed tomography (MSCT) in an ex vivo model. In three ex vivo specimens of left coronary arteries in oil, MSCT was performed after intracoronary injection of four solutions of contrast material (400 mgI/ml iomeprol). The four solutions were diluted as follows: 1%, 1/200, 1/80, and 1/20. All scans were performed with the following parameters: slices/collimation 16/0.75 mm, rotation time 375 ms. Each specimen was scored for the presence of atherosclerotic plaques. In each plaque the attenuation was measured in four regions of interest for lumen, plaque (non-calcified thickening of the vessel wall), calcium, and surrounding (oil surrounding the vessel). The results were compared with a one-way analysis of variance test and were correlated with Pearson's test. There were no significant differences in the attenuation of calcium and oil in the four solutions. The mean attenuation in the four solutions for lumen (35 +/- 10, 91 +/- 7, 246 +/- 18, 511 +/- 89 HU) and plaque (22 +/- 22, 50 +/- 26, 107 +/- 36, 152 +/- 67 HU) was significantly different between each decreasing dilution (p<0.001). The mean attenuation of lumen and plaque of coronary plaques showed high correlation, while the values were significantly different (r=0.73; p<0.001). Intracoronary attenuation modifies significantly the attenuation of plaques assessed with MSCT.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据