期刊
JOURNAL OF RHEOLOGY
卷 49, 期 4, 页码 919-922出版社
JOURNAL OF RHEOLOGY AMER INST PHYSICS
DOI: 10.1122/1.1942501
关键词
-
类别
When zeroing the gap between two parallel plates, the squeeze flow of air results in a detectable normal force at a finite height that is misinterpreted in rheometry as an indicator for the zero gap position. This leads to a substantial gap error using well-aligned parallel plates that is accurately predicted using Stefan's equation for squeeze flow. Minimizing or accounting for this error enables accurate rheological measurements to be performed in torsional flow using the parallel plate geometry at narrow gap heights, thus allowing shear rates of over 10(5) s(-1) to be achieved. (C) 2005 The Society of Rheology.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐