4.2 Article

Improving the performance of high-resolution X-ray spectrometers with position-sensitive pixel detectors

期刊

JOURNAL OF SYNCHROTRON RADIATION
卷 12, 期 -, 页码 467-472

出版社

BLACKWELL PUBLISHING
DOI: 10.1107/S0909049505010630

关键词

X-ray spectrometers; diced analyzer crystals; position-sensitive detectors

向作者/读者索取更多资源

A dispersion-compensation method to remove the cube-size effect from the resolution function of diced analyzer crystals using a position-sensitive two-dimensional pixel detector is presented. For demonstration, a resolution of 23 meV was achieved with a spectrometer based on a 1 m Rowland circle and a diced Si( 555) analyzer crystal in a near-backscattering geometry, with a Bragg angle of 88.5 degrees. In this geometry the spectrometer equipped with a traditional position-insensitive detector provides a resolution of 190 meV. The dispersion-compensation method thus allows a substantial increase in the resolving power without any loss of signal intensity.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据