Spin-echo MRS in humans at high field: LASER localisation using FOCI pulses

期刊

JOURNAL OF MAGNETIC RESONANCE
卷 175, 期 1, 页码 30-43

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jmr.2005.03.009

关键词

spin-echo; MRS; LASER; localisation; FOCI pulses

资金

  1. Wellcome Trust Funding Source: Medline

向作者/读者索取更多资源

Significant improvements in spin-echo MRS are possible when voxel localisation is performed using high bandwidth frequency offset corrected inversion (FOCI) pulses as opposed to more conventional lower bandwidth pulses. The reduced chemical shift displacement errors result in a spectrum that more accurately reflects the actual metabolite distribution within any region of interest that is selected graphically on a series of scout images, and can lead to improved metabolite detection in the case of homonuclear J-coupled spins. At 4.7 T, FOCI pulses with a 20 kHz bandwidth result in extremely sharp and uniform selection profiles, and negligible contamination from outside of the voxel of interest, for all signals in the H-1 spectral range that is normally studied. A 'FOCI' adiabatic half-passage is observed to provide good excitation over the 1 H spectral range. Single shot performance with echo-time (TE) >= 48 ins is reported using a four-port drive birdcage head coil. GAMMA simulations show that, for many detectable metabolites at 4.7 T, LASER localisation using FOCI pulses with TE = 48 ms results in H-1 anti-phase spectral components that are the same order as would be obtained from a symmetric PRESS sequence with TE = 32 ms. Timing schemes are proposed to enable good measurement of lactate with very little signal loss arising from chemical shift displacement errors at TE = 144 and 288 ms. (c) 2005 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据