期刊
JOURNAL OF MAGNETIC RESONANCE
卷 175, 期 2, 页码 275-284出版社
ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jmr.2005.04.014
关键词
small-volume EPR; planar microresonators; signal-to-noise ratio; sensitivity; microwave efficiency factor
EPR resonators on the basis of standing-wave cavities are optimised for large samples. For small samples it is possible to design different resonators that have much better power handling properties and higher sensitivity. Other parameters being equal, the sensitivity of the resonator can be increased by minimising its size and thus increasing the filling factor. Like in NMR, it is possible to use lumped elements; coils can confine the microwave field to volumes that are much smaller than the wavelength. We discuss the design and evaluation of EPR resonators on the basis of planar microcoils. Our test resonators, which operate at a frequency of 14 GHz, have excellent microwave efficiency factors, achieving 24 ns pi/2 EPR pulses with an input power of 17 mW. The sensitivity tests with DPPH samples resulted in the sensitivity value 2.3 x 10(9) spins (.) G(-1)Hz(-1/2) at 300 K. (c) 2005 Elsevier Inc. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐