4.5 Article

The analysis of thiamin and its derivatives in whole blood samples under high pH conditions of the mobile phase

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2005.05.027

关键词

thiamin; vitamin B1; HPLC; whole blood analysis; high pH mobile phase

向作者/读者索取更多资源

In this study a protocol for the analysis of thiamin and thiamin coenzymes in whole blood was developed. Thiamin and its coenzymes are analyzed by reversed phase liquid chromatography (RPLC), precolumn derivatisation with alkaline potassium ferricyanide and fluorescence detection, all at pH 10. Under these relatively high pH conditions the delectability of the analytes and the robustness of the method were substantially improved. The use of a high pH resistant RPLC column was a crucial step in developing this analysis method. Reproducibility, linearity, recovery, detection limit and column robustness were investigated. The within-batch CV was < 2.5%, the between-batch CV < 4.5%. The method was linear far above the physiological relevant concentration level. Recovery was almost 100% on an average. The limit of quantification was 1 nmol/l. The robustness of the RPLC column proved to be very high. Up to 1500 injections hardly any substantial changes in retention times and efficiency were observed. In summary: Using a high pH resistant RPLC column resulted in a robust, sensitive and precise method for the analysis of total Vitamin B I and especially of TDP. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据