4.7 Article Proceedings Paper

A novel preconcentration method for determination of iron and lead using Chromosorb-103 and flame atomic absorption spectrometry

期刊

ANALYTICA CHIMICA ACTA
卷 547, 期 1, 页码 26-30

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2005.01.056

关键词

syringe-connected minicolumn method; flame atomic absorption spectrometry; solid phase extraction; Chromosorb-103; bovine liver

向作者/读者索取更多资源

Syringe connected-minicolumn (SCM) method for the preconcentration/separation of iron and lead prior to their determination by flame atomic absorption spectrometry (FAAS) was developed. The proposed method is substituted for classical batch and column methods. The method proposed was compared with the column method with respect to easiness, rapidness, simplicity and some analytical performance criteria such as recovery, precision, accuracy and risk of contamination. A minicolumn was filled with Chromosorb-103 as a sorbent and connected with a syringe. The experimental conditions such as pH of sample, concentration and amount of eluent, flow rates of sample and eluent, concentration of 8-hydroxyquinoline (oxine) as a complexing agent were optimized. The sample solution with or without oxine was drawn into the syringe and discharged manually passing through the resin. Analytes were retained at pH >= 5 and eluted with I M nitric acid in acetone quantitatively. Analyte elements could be concentrated up to 200-fold. The iron and lead in river-water samples and in certified reference Bovine Liver (NIST 1577b) were quantitatively recovered with relative standard deviation lower than 10%. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据