4.4 Article

Polymer flip-chip bonding of pressure sensors on a flexible Kapton film for neonatal catheters

期刊

JOURNAL OF MICROMECHANICS AND MICROENGINEERING
卷 15, 期 9, 页码 1729-1735

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0960-1317/15/9/015

关键词

-

向作者/读者索取更多资源

In this paper, we describe the development of a new approach to mounting silicon pressure microsensors inside a 5 French (1.67 mm outer diameter) neonatal catheter tube for intravascular blood pressure measurements. Kapton film was used as a low-cost, flexible interface between the flip-chip bonded sensors at one end and wires to external electronics at the other end. Conductive polymer paste was used to flip-chip bond sensors to the flexible Kapton substrate at low temperature, while gold traces were used to form electrical interconnects on the film surface. Conductivity and bonding strength of the resulting polymer bumps were on the order of 100 m Omega and 4.43 mN mm(-2), respectively. Piezoresistive pressure microsensors were successfully mounted on Kapton carriers, packaged inside catheter tubes, and characterized in both gas and liquid environments. The sensors exhibited sensitivity of approximately 60 mu V mmHg(-1) over the target pressure range of 0-350 mmHg. Two-week tests in water at 27 degrees C and 36 degrees C showed only slight variations in contact resistances, indicating that the developed approach to packaging microsensors was sufficiently robust for the target application.