4.4 Article

On the routine use of soft X-rays in macromolecular crystallography. Part III. The optimal data-collection wavelength

期刊

出版社

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S0907444905021475

关键词

-

向作者/读者索取更多资源

Complete and highly redundant data sets were collected at different wavelengths between 0.80 and 2.65 angstrom for a total of ten different protein and DNA model systems. The magnitude of the anomalous signal-to-noise ratio as assessed by the quotient R-anom/R-r.i.m. was found to be influenced by the data-collection wavelength and the nature of the anomalously scattering substructure. By utilizing simple empirical correlations, for instance between the estimated Delta F/F and the expected R-anom or the data-collection wavelength and the expected R-r.i.m., the wavelength at which the highest anomalous signal-to-noise ratio can be expected could be estimated even before the experiment. Almost independent of the nature of the anomalously scattering substructure and provided that no elemental X-ray absorption edge is nearby, this optimal wavelength is 2.1 angstrom.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据