期刊
TALANTA
卷 67, 期 3, 页码 571-578出版社
ELSEVIER
DOI: 10.1016/j.talanta.2005.03.018
关键词
CDM-PSS DGT; speciation; trace metals; waters
The speciation measurements of trace metals by the diffusive gradients in thin-films technique (DGT) using a poly(4-styrenesulfonate) (PSS) aqueous solution as a binding phase and a cellulose dialysis membrane (CDM) as a diffusive layer, CDM-PSS DGT, were investigated and showed good agreement with computer modelling calculations. The diffusion coefficients of ethylenediaminetetraacetic acid (EDTA) complexes with Cd2+ and Cu2+ were measured and compared with those of the inorganic metal ions. CDM-PSS DGT device was tested for speciation measurement in sample solutions containing EDTA, tannic acid (TA), glucose (GL), dodecylbenzenesulfonic acid (DBS) and humic acid (HA) as complexing ligands forming organic complexes with varying stability constants. Lower percentages of DGT labile copper concentrations over total filterable copper concentrations obtained from the deployments in freshwater sites indicated that copper complexes with organic matter were basically not measured by the devices. (c) 2005 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐