A rapid and sensitive HPLC method using a monolithic column has been developed for quantification of famotidine in plasma. The assay enables the measurement of famotidine for therapeutic drug monitoring with a minimum detectable limit of 5 ng ml-(1). The method involves simple, one-step extraction procedure and analytical recovery was complete. The separation was carried out in reversed-phase conditions using a Chromolith Performance (RP-18e, 100 mm x 4.6 mm) column with an isocratic mobile phase consisting of 0.03 M disodium hydrogen phosphate buffer-acetonitrile (93:7, v/v) adjusted to pH 6.5. The wavelength was set at 267 nm. The calibration curve was linear over the concentration range 20-400 ng ml(-1). The coefficients of variation for inter-day and intra-day assay were found to be less than 8%. (c) 2005 Elsevier B.V. All rights reserved.