4.6 Article

Sweeping-window arc therapy: an implementation of rotational IMRT with automatic beam-weight calculation

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 50, 期 18, 页码 4317-4336

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/50/18/006

关键词

-

向作者/读者索取更多资源

Sweeping-window arc therapy (SWAT) is a variation of intensity-modulated radiation therapy (IMRT) with direct aperture optimization (DAO) that is initialized with a leaf sequence of sweeping windows that move back and forth periodically across the target as the gantry rotates. This initial sequence induces modulation in the dose and is assumed to be near enough to a minimum to allow successful optimization, done with simulated annealing, without requiring excessive leaf speeds. Optimal beam weights are calculated analytically, with easy extension to allow for variable beam weights. In this paper SWAT is tested on a phantom model and clinical prostate case. For the phantom, constant and variable beam weights are used. Although further work (in particular, improving the dose model) is required, the results show SWAT to be a feasible approach to generating deliverable dynamic arc treatments that are optimized.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据