3.8 Article

Corrosion behaviour, metal release and biocompatibility of implant materials coated by TiO2-sol gel chemistry

期刊

BIOMEDIZINISCHE TECHNIK
卷 50, 期 10, 页码 320-329

出版社

FACHVERLAG SCHIELE SCHON
DOI: 10.1515/BMT.2005.045

关键词

TiO2-coating; Ti6Al4V-and Co28Cr6Mo-alloys; ion release; in vitro cell test

向作者/读者索取更多资源

Alloys based on titanium or cobalt have been used as implant materials for decades with good success. Because of their natural oxide layer these alloys reveal good corrosion behaviour. In contact with physiological solution metal release takes place, which can cause inflammation. Coatings can improve the corrosion behaviour In this study Ti6Al4V and Co28Cr6Mo alloys, which are frequently used as implant materials, were tested. Polished discs of these alloys and polished discs, which were coated with TiO2-layers by sol-gel chemistry, were compared regarding their corrosion behaviour and metal ion releasing. The releasing of Al, V, Ti, Co, Cr and Mo was quantified by ICP-MS analysis. The TiO2-coating reduced the release of all ions except of the Al-ion. Both alloys showed a deviating kinetic of ion releasing. In addition, cell response (cell vitality, cell proliferation, endothelial marker CD31 and actin allocation) of osteoblasts and endothelial cells were investigated.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据