4.5 Article

MALDI-MS imaging of features smaller than the size of the laser beam

期刊

出版社

SPRINGER
DOI: 10.1016/j.jasms.2005.06.006

关键词

-

资金

  1. NIAAA NIH HHS [1F32 AA014459] Funding Source: Medline
  2. NIDA NIH HHS [P30 DA018310, DA14879] Funding Source: Medline

向作者/读者索取更多资源

The feasibility of matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) imaging of features smaller than the laser beam size has been demonstrated. The method involves the complete ablation of the MALDI matrix coating the sample at each sample position and moving the sample target a distance less than the diameter of the laser beam before repeating the process. In the limit of complete sample ablation, acquiring signal from adjacent positions spaced by distances smaller than the sample probe enhances image resolution as the measured analyte signal only arises from the overlap of the laser beam size and the non-ablated sample surface. Image acquisition of features smaller than the laser beam size has been demonstrated with peptide standards deposited on electron microscopy calibration grids and with neuropeptides originating from single cells. The presented MS imaging technique enables similar to 25 mu m imaging spatial resolution using commercial MALDI mass spectrometers having irregular laser beam sizes of several hundred micron diameters. With appropriate sampling, the size of the laser beam is not a strict barrier to the attainable MALDI-MS imaging resolution.