4.3 Article Proceedings Paper

New results for single stage low energy carbon AMS

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.nimb.2005.06.168

关键词

accelerator mass spectrometry; background; C-14 (or carbon-14); precision; single stage AMS

向作者/读者索取更多资源

A new configuration of the NEC single stage, low energy carbon AMS system (U.S. Patent 6,815,666 B2) has been built and tested. The injector includes two 40-sample ion sources, electrostatic and magnetic analysis, and fast sequential injection. The gas stripper, analyzing magnet, electrostatic analyzer, and detector are on an open air 250 kV deck. Both C-12 and C-13 currents are measured on the deck after the stripper, and an SSB detector is used for C-14 counting. Injected C-12 and mass 13 (C-13 and (CH)-C-12) currents are also measured. Automated controls follow a user-specified run list for unattended operation. Initial test results show precision for C-14/C-12 ratios of better than 5 per mil, and backgrounds for unprocessed graphite of less than 0.005 x modern. We will report final results for precision, background, and throughput and discuss related design features. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据