4.5 Article

Impact of central venous catheter type and methods on catheter-related colonization and bacteraemia

期刊

JOURNAL OF HOSPITAL INFECTION
卷 61, 期 2, 页码 139-145

出版社

W B SAUNDERS CO LTD
DOI: 10.1016/j.jhin.2005.02.012

关键词

bacterial infections/prevention and control; catheterization central venous/adverse effects; crossinfection/prevention and control; Bacteraemia

向作者/读者索取更多资源

A prospective, randomized, controlled, multi-centre clinical. trial was performed to test the effectiveness of an antimicrobial central venous catheter (CVC) made of polyurethane integrated with silver, platinum and carbon black (Vantex (R)). Adults expected to require a CVC for more than 60 h were eligible, and were randomized to receive the test or control catheter. All CVCs were inserted with new venipunctures using full aseptic technique. Following catheter removal, the distal. tip and an intracutaneous segment were removed and cultured using semi quantitative and quantitative methods. Peripheral blood samples were obtained and cultured to confirm cases of catheter- related bloodstream infection (CRBSI). Bacterial and fungal organisms were identified by standard microbiological methods. Catheter placement was performed primarily in the intensive care unit (50%) or operating theatre (42%). Complete data could be evaluated for 539 patients (77%). The mean duration of CVC placement was 149.3 h (six days). There were no significant differences in colonization or bacteraemia rates between the test and control catheters. The overall colonization rate wits not particularly low (24.5%), and yet CVC-related bacteraemia occurred in only 1.4% of patients, and CRBSI occurred in only one patient from the control. group (0.2%). Insertion site and dressing change frequency were significantly associated with the colonization rate. Although CVCs with antimicrobial features have been associated with a decrease in catheter-related colonization and bacteraemia, this study demonstrated that infection rates may depend more on non -catheter- related factors, such as adherence to infection control standards, selection of insertion site, duration of CVC placement, and dressing change frequency. As microbial resistance increases, clinicians should make maximal use of these processes to reduce catheter- related infections. (c) 2005 The Hospital Infection Society. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据