Development and evaluation of a reference measurement procedure for the determination of estradiol-17β in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry

期刊

ANALYTICAL CHEMISTRY
卷 77, 期 19, 页码 6359-6363

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac050837i

关键词

-

向作者/读者索取更多资源

Estradiol is the most potent natural estrogen and is derived from the ovaries. Its concentration in blood is measured to determine ovarian function. A reference measurement procedure for estradiol in serum involving isotope-dilution coupled with liquid chromatography-tandem mass spectrometry (LC/MS/MS) has been developed and critically evaluated. A deuterated estradiol (estradiol-d(3)) was used as an internal standard. The estradiol and its internal standard were extracted from serum matrix using solid-phase extractions and derivatized with dansyl chloride prior to reversed-phase LC/MS/MS. The accuracy of the measurement was evaluated by a comparison of results of this reference method on lyophilized human serum reference materials for estradiol [Certified Reference Materials (CRMs) 576, 577, and 578] with the certified values determined by gas chromatography/mass spectrometry (GC/MS) reference methods and by a recovery study for the added estradiol. The results of this method for estradiol agreed with the certified values within the uncertainty of the measurements for the three CRMs. The recovery of the added estradiol ranged from 100.7 to 101.8%. This method was applied to the determination of estradiol in frozen serum samples from three individual female donors. Excellent reproducibility was obtained with within-set coefficient of variations (CVs) ranging from 0.6 to 2.2% and between-set CVs ranging from 0.2 to 0.4%. Excellent linearity was also obtained, with correlation coefficients of all linear regression lines (measured intensity ratios vs mass ratios) ranging from 0.998 to 1.000. The detection limit at a signal-to-noise ratio of similar to 3 was 0.6 pg of estradiol (or 1 ng/L, as expressed as a concentration). This well-characterized LC/MS/MS method for serum estradiol, which demonstrates good accuracy and precision, low susceptibility to interferences, and comparability with GC/MS reference methods, qualifies as a reference measurement procedure that can be used to provide an accuracy base to which routine methods for estradiol can be compared and that will serve as a standard of higher order for measurement traceability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据