4.4 Review

Response of brain tissue to chronically implanted neural electrodes

期刊

JOURNAL OF NEUROSCIENCE METHODS
卷 148, 期 1, 页码 1-18

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jneumeth.2005.08.015

关键词

biocompatibility; chronic; recording electrode array; cortical implant; glial scar; neuroprosthetic; wound healing; foreign body reaction; neural probe

向作者/读者索取更多资源

Chronically implanted recording electrode arrays linked to prosthetics have the potential to make positive impacts on patients suffering from full or partial paralysis. Such arrays are implanted into the patient's cortical tissue and record extracellular potentials from nearby neurons, allowing the information encoded by the neuronal discharges to control external devices. While such systems perform well during acute recordings, they often fail to function reliably in clinically relevant chronic settings. Available evidence suggests that a major failure mode of electrode arrays is the brain tissue reaction against these implants, making the biocompatibility of implanted electrodes a primary concern in device design. This review presents the biological components and time course of the acute and chronic tissue reaction in brain tissue, analyses the brain tissue response of current electrode systems, and comments on the various material science and bioactive strategies undertaken by electrode designers to enhance electrode performance. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据