4.7 Article

Automated spectrophotometric analysis of mitochondrial respiratory chain complex enzyme activities in cultured skin fibroblasts

期刊

CLINICAL CHEMISTRY
卷 51, 期 11, 页码 2110-2116

出版社

OXFORD UNIV PRESS INC
DOI: 10.1373/clinchem.2005.050146

关键词

-

向作者/读者索取更多资源

Background: Mitochondrial respiratory chain complex (RCC) disorders may occur as commonly as 1 in 8500 individuals. Because of the great variability of phenotypic presentations, measurement of individual RCC enzyme activities is a crucial diagnostic process. Current assay methods are time-consuming and labor-intensive and thus constitute a major impediment to clinical practice. A method with a faster turnaround time would therefore be beneficial. Method: We developed an automated spectrophotometric method. for measuring the respiratory chain enzyme activities of complex I, complex II + III, and complex IV with the Hitachi 912, an automated spectrophotometer. Mitochondrial citrate synthase was also determined for normalization of the RCC activities. Results: A blinded method comparison with samples from an external testing center yielded a 91% concordance of interpretations. Mean intraassay imprecision (as CV; n = 20) in a single batch analysis of each RCC was 5.9%. Interassay imprecision, evaluated on 2 samples harvested and analyzed 3 times each, gave mean CVs of 10%-18%. Conclusions: With this automated method, a panel of RCC enzyme activities can be determined in < 2 h. In addition, an immunoblot assay using monoclonal antibodies against specific subunits of RCC enzyme complexes can be informative in cases of borderline enzyme activity. Our results suggest that in vitro diagnosis of RCC enzyme deficiencies in skin fibroblasts is an effective alternative to invasive muscle biopsy. 2005 American Association for Clinical Chemistry.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据