Prospective correction of affine motion for arbitrary MR sequences on a clinical scanner

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 54, 期 5, 页码 1130-1138

出版社

WILEY
DOI: 10.1002/mrm.20686

关键词

coronary MR angiography; RF excitation; respiratory motion; prospective motion correction; affine motion model; model calibration

向作者/读者索取更多资源

The concept of prospective 3D affine motion correction was generalized, based on the Bloch equations, for signal excitation and sampling using arbitrary MR sequences. The technique was implemented on a clinical MRI scanner for Cartesian, radial, and spiral imaging sequences, as well as for 2D spatially selective RF excitation pulses. A patient-specific motion model steered by real-time navigators was employed to account for the additional degrees of freedom provided by the affine motion model. Different navigator concepts (multiple spatial and temporal navigators, quadratic navigators and other motion sensors) were investigated, with the aim of improving the correlation between navigator information and the motion model. Experiments on moving phantoms are presented to prove the technical feasibility of the approach. In vivo experiments on coronary MRA and renal MRI show the potential of the method for cardiac and abdominal applications hampered by respiratory motion. Magn Reson Med 54:1130-1138, 2005. (c) 2005 Wiley-Liss, Inc.