4.5 Article

Two-color pump-probe laser spectroscopy instrument with picosecond time-resolved electronic delay and extended scan range

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 76, 期 11, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.2126808

关键词

-

向作者/读者索取更多资源

An electronically delayed two-color pump-probe instrument was developed using two synchronized laser systems. The instrument has picosecond time resolution and can perform scans over hundreds of nanoseconds without the beam divergence and walk-off effects that occur using standard spatial delay systems. A unique picosecond Ti: sapphire regenerative amplifier was also constructed without the need for pulse stretching and compressing optics. The picosecond regenerative amplifier has a broad wavelength tuning range, which suggests that it will make a significant contribution to two-color pump-probe experiments. To test this instrument we studied the rotational correlation relaxation of myoglobin (tau(r)= 8.2 +/- 0.5 ns) in water as well as the geminate rebinding kinetics of oxygen to myoglobin (k(g1) = 1.7 x 10(11) s(-1), k(g2) = 3.4 x 10(7) s(-1)). The results are consistent with, and improve upon, previous studies. (c) 2005 American Institute of Physics.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据