4.6 Article

Optimized transition-edge x-ray microcalorimeter with 2.4 eV energy resolution at 5.9 keV

期刊

APPLIED PHYSICS LETTERS
卷 87, 期 19, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.2061865

关键词

-

向作者/读者索取更多资源

We present measurements from a series of transition-edge x-ray microcalorimeters designed for optimal energy resolution. We used the geometry of the sensors to control their heat capacity and employed additional normal metal features and a perpendicular magnetic field to control the sharpness of the superconducting-to-normal transition. These degrees of control allow an optimal selection of sensor saturation energy and noise. Successive design changes improved the measured energy resolution of the sensors from 4.5 eV full width at half maximum at 5.9 keV to 2.4 eV at 5.9 keV. Sensors with this energy resolution are well matched to applications in x-ray astrophysics and terrestrial materials analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据