4.7 Article Proceedings Paper

Analytical performance and characterization of MPA-Gly-Gly-His modified sensors

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 111, 期 -, 页码 540-548

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2005.03.046

关键词

biosensors; peptides; Gly-Gly-His; copper; electrochemistry

向作者/读者索取更多资源

This paper studies the analytical performance and characterization of a peptide modified biosensor towards detection of CU2+ ions. The peptide modified biosensor is developed by carbodiimide attachment of the Gly-Gly-His tripeptide onto a 3-mercaptopropionic acid (MPA) modified gold surface. X-ray photoelectron spectroscopy (XPS) was used to characterize the steps in the biosensor fabrication. Detection of CU2+ ions are determined by cyclic and Osteryoung square wave voltammetry. The analytical performance of this biosensor is found to be affected by the pH, salt concentration, temperature and accumulation time of the test solution. Interference studies and investigations of the reusability of this biosensor have also been performed. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据