4.3 Article

Loop-mediated isothermal amplification for the rapid detection of Salmonella

期刊

FEMS MICROBIOLOGY LETTERS
卷 253, 期 1, 页码 155-161

出版社

OXFORD UNIV PRESS
DOI: 10.1016/j.femsle.2005.09.032

关键词

Salmonella; LAMP; rapid; sensitive; detection

向作者/读者索取更多资源

Loop-mediated isothermal amplification (LAMP) assay detected Salmonella within 60 min. The 220 strains of 39 serotypes of Salmonella subsp. enterica and 7 strains of Salmonella enterica subsp. arizonae were amplified, but not 62 strains of 23 bacterial species other than Salmonella. The sensitivity of the LAMP assay was found to be >2.2 cfu/test tube using nine serotypes. The specificity was similar to that of a PCR assay, but the sensitivity of LAMP was greater. Both fluorescence and turbidity were able to detect the products in the LAMP assay. S. enteritidis in a liquid egg sample artificially inoculated with the organism was detected by the LAMP assay at 2.8 cfu/test tube, although negative by PCR assay. These results indicate that the LAMP assay is a rapid, specific and sensitive detection method for Salmonella. (c) 2005 Federation of European Microbiological Societies. Published by Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据