4.5 Article

Design and performance of a thermal-neutron double-crystal diffractometer for USANS at NIST

期刊

JOURNAL OF APPLIED CRYSTALLOGRAPHY
卷 38, 期 -, 页码 1004-1011

出版社

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S0021889805032103

关键词

-

向作者/读者索取更多资源

An ultra-high-resolution small-angle neutron scattering (USANS) double-crystal diffractometer (DCD) is now in operation at the NIST Center for Neutron Research (NCNR). The instrument uses multiple reflections from large silicon (220) perfect single crystals, before and after the sample, to produce both high beam intensity and a low instrument background suitable for small-angle scattering measurements. The minimum detector background to beam intensity ratio (noise-to-signal, N/S) for q >= 5 x 10(-4) angstrom(-1) is 4 x 10(-7). The instrument uses 2.38 A wavelength neutrons on a dedicated thermal neutron beam port, producing a peak flux on the sample of 17300 cm(-2) s(-1). The typical measurement range of the instrument extends from 3 x 10(-5) angstrom(-1) to 5 x 10(-3) angstrom(-1) in scattering wavevector (q), providing information on material structure over the size range from 0.1 mm to 20 mm. This paper describes the design and characteristics of the instrument, the mode of operation, and presents data that demonstrate the instrument's performance.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据