Accuracy of implantation of components in the Oxford knee using the minimally invasive approach

期刊

KNEE
卷 12, 期 6, 页码 405-409

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.knee.2005.03.003

关键词

Oxford knee; accuracy; minimally invasive

向作者/读者索取更多资源

Screened postoperative X-rays of 224 Oxford knees implanted through the minimally invasive approach were analysed using 16 criteria. The technique was as recommended by the Oxford Group except that the femoral intramedullary rod was used only as a guide to flexion/extension of the femoral component. All femoral components were within the recommended range for varus/valgus and mediolateral position. Eighteen femoral components were either too flexed or extended, but by a maximum of only 10 degrees. Tibial components were inserted in slight varus (mean 1.8 degrees, S.D. 8.8, range 10 to -10). All were within the range for posterior slope in spite of difficulty in establishing objective landmarks. Tibial coverage was imprecise with the phase 3 implants which were short and broad relative to the cut surface. In 36% of knees the tibial tray was implanted too anterior resulting in posterior under hang. There have been no clinical signs to indicate tipping of the meniscus in deep flexion. The depth of cement penetration under the keel was excessive in three cases, resulting from porotic bone rather than technical error in the depth of the cut. There has been one case of femoral component loosening following a fall and no tibial loosening in spite of imprecise alignment of some components. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据